Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8: (1 year) CAL VER /5520 INSTRUMENT: Fluke 10: (1 year) CAL VER /5520 DATE: 22-Sep-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 26 NUMBER OF LINES: 108 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # Fluke 10 Series Service Manual (PN 900824 March 1992, Rev. 1, 7/93). # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F 1.002 ASK+ K 1.003 HEAD EQUIPMENT SETUP 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP No warm-up time requirement. 1.005 DISP Ambient temperature: 18C - 28C. 1.005 DISP Relative humidity: less than 90%. 1.006 DISP Move switch to OFF position. 1.006 DISP 1.006 DISP Note: Erroneous result may occur if switch is not 1.006 DISP placed in OFF position before beginning tests. 1.007 HEAD {RESISTANCE TESTS} 1.008 DISP Move switch to the right position. 1.008 DISP Press SELECT for Ohms 1.009 HEAD RESISTANCE TESTS: {400 Ohm Range} 1.010 5520 400 0.0Z +0.2U 2W 2.001 5520 400 1.0Z 0.2U 2W 3.001 5520 400 350.0Z 5.5U 2W 4.001 HEAD RESISTANCE TESTS: {4 kOhm Range} 4.002 5520 4 1.000kZ 0.016U 2W 5.001 HEAD RESISTANCE TESTS: {40 kOhm Range} 5.002 5520 40 35.00kZ 0.54U 2W 6.001 HEAD RESISTANCE TESTS: {400 kOhm Range} 6.002 5520 400 100.0kZ 1.6U 2W 7.001 HEAD RESISTANCE TESTS: {4 MOhm Range} 7.002 5520 4 2.900MZ 0.045U 2W 8.001 HEAD RESISTANCE TESTS: {40 MOhm Range} 8.002 5520 40 35.00MZ 0.70U 2W 9.001 HEAD {CONTINUITY TEST} 9.002 DISP Press SELECT for Continuity. 9.003 5520 2.000V 0.032U 2W 10.001 HEAD {AC VOLTAGE TESTS} 10.002 DISP Move switch to the center position. 10.002 DISP 10.002 DISP Press SELECT for VAC. 10.003 HEAD AC VOLTAGE TESTS: {4V Range} 10.004 5520 0.000V S 2W 10.005 MEMI Enter UUT reading in volts AC: 10.006 MEME 10.007 MEMC 4 V +0.003U 11.001 5520 4 0.010V 0.003U 50H SI 2W 12.001 5520 4 0.010V 0.003U 400H SI 2W 13.001 5520 4 3.500V 0.105U 50H SI 2W 14.001 5520 4 3.500V 0.105U 400H SI 2W 15.001 HEAD AC VOLTAGE TESTS: {40V Range} 15.002 5520 40 35.00V 1.05U 50H SI 2W 16.001 5520 40 35.00V 1.05U 400H SI 2W 17.001 HEAD AC VOLTAGE TESTS: {400V Range} 17.002 5520 400 350.0V 10.5U 80H SI 2W 18.001 HEAD AC VOLTAGE TESTS: {600V Range} 18.002 5520 600 600V 20U 400H SI 2W 19.001 HEAD {DC VOLTAGE TESTS} 19.002 DISP Press SELECT for VDC. 19.003 HEAD DC VOLTAGE TESTS: {4V Range} 19.004 DISP Press RANGE to select the 4V Range. 19.005 5520 4 0.000V 0.002U 2W 20.001 HEAD DC VOLTAGE TESTS: {40V Range} 20.002 DISP Press RANGE to select the 40V Range. 20.003 5520 40 0.02V 0.01U 2W 21.001 HEAD DC VOLTAGE TESTS: {400V Range} 21.002 DISP Press RANGE to select the 400V Range. 21.003 5520 400 -0.2V 0.1U 2W 22.001 HEAD DC VOLTAGE TESTS: {4V Range} 22.002 DISP Press RANGE for 2s to select autorange. 22.003 5520 4 3.500V 0.055U 2W 23.001 HEAD DC VOLTAGE TESTS: {40V Range} 23.002 5520 40 -35.00V 0.54U 2W 24.001 HEAD DC VOLTAGE TESTS: {400V Range} 24.002 5520 400 100.0V 1.6U 2W 25.001 HEAD DC VOLTAGE TESTS: {600V Range} 25.002 5520 600 -600V 10U 2W 26.001 END